| বিশেষভাবে তুলে ধরা: | KNF diaphragm vacuum pump replacement,BAXIT gas sampling pump,N89KNE vacuum pump for DSC |
||
|---|---|---|---|
BAXIT replace KNF Diaphragm Vacuum pump N89KNE N89KTE Gas Sampling pump
Adapt to the environment:
This product is used in industrial process gas continuous analysis system (extraction measurement), for gas extraction, internal anti-corrosion treatment, especially suitable for flue gas analysis. Extraction of gases such as SO2, NOX, CO2, CO, N2, O2, or a mixer can be performed.
![]()
Technical Parameters:
| Model | N89KNE | N89KTE |
| Pump head | PPS | PPS |
| Diaphragm | EPDM | PTFE-coated |
| Valves | CR | FFPM |
| Flow rate at atm. pressure (l/min) | 9.5 | 9.5 |
| Ultimate vacuum (mbar abs.) | 100 | 170 |
| Max. operating pressure (bar) | 0.5 | 0.5 |
| Permissible ambient temperature (ºC) | "+5 ... +40 | "+5 ... +40 |
| Permissible media temperature (ºC) | "+5 ... +40 | "+5 ... +40 |
| Voltage (V) | 230 | 230 |
| Motor | Shaded pole motor | Shaded pole motor |
| Protection class motor | IP 00 | IP 00 |
| Frequency (Hz) | 50 | 50 |
| Power P1 (W) | 60 | 60 |
| Imax (A) | 0.6 | 0.6 |
| Weight (kg) | 1.5 | 1.5 |
Installation method:
This product is panel mounted, the mounting hole spacing is: 48mm, 48mm. The mounting hole size is: 4-Φ5.
Daily maintenance:
This product is an oil-free diaphragm vacuum pump, no need to refuel. If the inside is clogged or contaminated, open the top four screws, clean the gasket, and clean the pump membrane. If there is serious pollution, resulting in air leakage or insufficient air supply during the sampling process, the pump membrane should be replaced in time.